C009 3H replaceable membrane electrolyzer
Technical
features:
1. The cell has a salt bridge (Lujin capillary)
with adjustable angle, which makes the reference electrode align with the
corrosion sample by adjusting the angle, which can reduce the R drop more
effectively
2. The electrolyzer should avoid chloride ion in corrosion
laboratory, so the solution filled in salt bridge is saturated potassium
nitrate solution.
3. The working electrode of the electrolytic cell
is equipped with sample support electrode, which can be replaced according to
your own requirements. The sample diameter is 14mm and the fixed area is 1cm
4. Configuration: electrolytic cell body, F-type
air inlet, salt bridge, liquid seal, PTFE plug sample support, reference
electrode, graphite rod electrode
5. The material of the cell is borosilicate glass
6. The cell is a relatively sealed cell system,
which is commonly used in three electrode system. The common volume of the cell
is 500ml and 1000m, and can be customized according to the requirements
7. The electrolyzer is equipped with F-type air
inlet which can ventilate up and down the liquid surface. According to the
different experiments, the atmosphere protection should be carried out. When
the experiment starts, please stop the ventilation under the liquid level and
only keep the ventilation on the liquid surface
8. Scope of application: it can be used to test and
determine metal samples in corrosive environment, which can ensure the rapid
and consistent test of a group of metal samples in liquid environment. It is
suitable for most electrochemical instruments used for corrosion test, and can
be used at room temperature and high temperature in corrosive environment
(except HF).